113

Test CODE NAME

113

Bilirubin, Direct

Methodology: Quantitative Spectrophotometry
Performed: Sun-Sat
Reported: 1-2 days
Specimen Required: Collect: Serum Separator Tube (SST) - 0.5 ml


Specimen preparation: Serum free from hemolysis is the recommended specimens. Allow blood samples to clot (15 mins). Separate the serum from the cells by centrifuging for 10 minutes. Store serum at 2-8°C until analysis. 
    
Storage/Transport: Refrigerated

Stability: Exposure to direct sunlight can decrease bilirubin in samples by 50% within one hour. When well protected from light, bilirubin in serum is stable for up to 3 days at 2 - 8°C and for approximately three months when stored at <-20°C in the dark. 
At Accu Reference: Refrigerated - 3 days: Frozen- 1 month

NOTE: * This test is approved for all states. *

Performing Lab
Accu Reference Medical Laboratory
CPT Code(s): 83021